Beckman Coulter DSC 700 AU Analyser Results into an SQL Server Table
- Data
- Result records from a Beckman Coulter DSC 700 AU biochemistry analyser: STX, a fixed-position record, ETX, a check character
- Industry
- Clinical laboratory
- Country
- Australia
- Interface
- RS-232
- Product
- Advanced Serial Data Logger, ASCII data parser and SQL Database Professional modules
- Result
- One row per record with system, rack, cup, sample type and number, specimen ID, job number and up to 20 test slot, result and flag columns
The task
A pathology lab in Western Australia already logged the serial output of its Beckman Coulter DSC 700 AU analyser with Advanced Serial Data Logger. The raw log was there; what the lab needed was each result record in an SQL Server table, with a column for every field of the record. The sample log showed the control characters the analyser uses and a quality-control run where the job number field carries a material name such as BIORAD1B instead of a number.
What the analyser sends
Every record is framed: STX (0x02), the record text, ETX (0x03) and one check character. A batch starts with a DB00 record followed by ENQ, ends with DE00, and the result records in between begin with D. An excerpt with the control characters shown as names (the quality-control record from the sample, results shortened):
<STX>DB00<ETX><ENQ> <STX>D 00001501 0002 BIORAD1B E001000000029 003000000103 005000000026 ...<ETX>7 <STX>DE00<ETX>
The record has a fixed layout. The header occupies the first 39 columns, column 40 holds the letter E, and from column 41 the results follow in groups of 20 characters: a three-digit test number, a nine-character value and a one-character flag.
column 1 D record type column 3 00 system number column 5 0015 rack number column 9 01 cup number column 11 (space) sample type column 12 0002 sample number column 3 00001501 specimen ID (system, rack and cup together) column 24 BIORAD1B job number, 14 characters, or the control material name column 40 E start of results column 41 001 test number of result 1 column 44 000000029 value of result 1 column 54 (space) flag of result 1 column 61 003 test number of result 2, and so on every 20 columns
In the sample every record appears twice, a few seconds apart. If your log shows the same, give the table a unique key on specimen ID, job number and the record time, or deduplicate on insert.
The parser configuration
The ASCII data parser module does the work with four settings:
- Data packet start signature
#02, end signature#03. The check character after ETX is dropped with the next frame's prefix. - A filter with the regular expression
D \d{2}, so that only result records are parsed and theDB00andDE00block markers are ignored. - Variables by fixed position, with the column and length from the table above. The job number is a text variable because of the control material names.
- Twenty result groups, each a slot, a value and a flag variable, named SLOT001 to FLAG020.
The configuration is here as the files that the parser's Load/Save button accepts: beckman-au-ascii-parser.ini (signatures, filter and the first three result groups) and beckman-au-parser-fields-20-results.txt (the full field list with 20 result groups, which you can paste into the variable list or extend to the number of tests your analyser reports). The same approach, fixed-position fields inside a framed record, is shown step by step in the PBX system data logger tutorial.
Into SQL Server
The SQL Database Professional module writes one row per parsed record through the native SQL Server driver. Create the table with one column per variable, text for the job number and the flags, numeric for the values, and map the variables to the columns in the module's settings. The result is the layout the lab asked for: a wide table whose row matches the incoming record. If you would rather have one row per test, the same data can be exported that way with a different variable layout; write to support with your table definition.
Try it yourself
The trial version of Advanced Serial Data Logger has every option described above. Configure it the same way and write to support if your device sends something different.
Related topics: Advanced Serial Data Logger
hereSerial port interface | RS232 port logger | Com port logger | Data acquisition | RS232 data logger | Data Logger Software For Electricity Meter | Replace WinWedge | Replace TWedge | RS232 to SQL database | Barcode to Excel